Operational Principles of LECO Carbon Sulfur Analyzer and Tungsten Granule

As a new and high-tech level analyzer, carbon sulfur analyzer is widely applied in analyzing the content of carbon and sulfur of the alloy, cast-iron, lime and other solid and fluid materials. It has a high accuracy and a wide measurement range when analyze carbon and sulfur. The modular and integrative design of the whole structure of the analyzer achieves an easier operation, a faster analysis, a greater precision of analysis, a more reliable working of instrument. LECO carbon sulfur analyzer has been kept a leading edge in its technology, and has a number of technology patents, which has become the most well-known brands of analytical instruments in the world.

Operational principles of LECO carbon sulfur analyzer and tungsten granule mainly include the operation procedure before starting, the opening of host, the entering of the main program, dank correction and the shutdown operation. First of all, startup the UPS equipment, the power supply is checked to ensure whether it can be used normally or not. Then turn off the air switch and the stabilized voltage supply, check whether the output stabilized voltage supply is 220V or not. Under normal circumstances of the input of power, then get into the boot process. Next, open the air supply of the carrier gas O2 and power gas Ar, and adjust it into the appropriate pressure.

The carrier gas O2 is 0.26MPa, power gas Ar is 0.26MPa. Firstly, turn off of power switch of the furnace equipment (switch is located in the front of lower part of the stove, a black switch), then turn off the power switch of the instrument detector. You can see that some electronic tube is lit in the furnace crucible, the flowmeter of the float should be placed at 3L / min. The mainframe computer and monitor power is turned on while the printer power and balance power is turned on. Since the balance is in balance and it display 0.0000, double-clicking the CS230 icon on your desktop, entering CS230 analysis process.

After the analyzer and computer communication can be carried out normally, checking the air leak condition, then enter the sub-menu to view corresponding parameters and graphics, the next step is to be tested after a stable of data in detection pool. Finally, turn off the software first before shutting down, then turn off the switch and the analyzer switch in the front of stove, then turn off the gas source, and the power.

tungsten granule

 

 

WeChat